FDA Recall Terminated

UniCel DxC 800 Part Number: A10411, Beckman Coulter, Inc.

Recall: Z-1100-2008 · Initiated April 5, 2007

Recall

Recall Number
Z-1100-2008
Event Number
45960
Firm
Beckman Coulter Inc
FEI Number
2050012
Product Code
JJE
Status
Terminated
Root Cause
Software design
Initiated
April 5, 2007
Posted
May 24, 2008
Terminated
January 15, 2013
Address
200 S Kraemer Blvd, Brea, CA, 92822-6208

Description

UniCel DxC 800 Part Number: A10411, Beckman Coulter, Inc.

Reason

Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors. -If a stall occurs during a GLUm test near the time of sample inject, results may be affected. -Affected GLUm results may be high or low.

Action

GLUm stir motors greater than one year old were replaced. Replacements began April 5, 2007. Based on the Risk Assessment, BCI decided that Customer Notification is not needed. GLUm stir motors greater than one year old will be replaced on high volume LX/DxC systems by 3rd quarter of 2007. Replacements began April 5, 2007. A new stirrer design is being developed and is expected to be available before 4th quarter of 2007. Implementation Strategy of Stock Control: 1. A Stop Ship was not necessary based on risk assessment. The possibility of stirrer motor failure may increase with time and duty cycles. Systems running a high volume of tests may have a greater likelihood of exhibiting stirrer motor failure with motors greater than one year old. 2. New Customers were placed on a preventative maintenance plan to have their motor replaced at one year cycle. Completed 3. A Technical Update was issued to notify Hotline and Field Service Personnel of the issues.

Distribution

Worldwide Distribution - USA including Canada.

Quantity

46 in the US; 1 in Canada