FDA Recall Terminated

Malecot Nephrostomy Catheter System, Product Number: M001224110, GTIN: 08714729030980 The Malecot and the Stenting Malecot Catheters are designed for percutaneous drainage within the renal collecting system while maintaining external access. The Stenting Malecot Nephrostomy Catheter with its extended tip is also designed for internal drainage from the ureteropelvic junction to the bladder.

Recall: Z-1099-2018 · Initiated December 12, 2017

Recall

Recall Number
Z-1099-2018
Event Number
79424
Firm
Boston Scientific Corporation
FEI Number
3005099803
Product Code
GBW
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
December 12, 2017
Terminated
October 24, 2019
Address
100 Boston Scientific Way, Marlborough, MA, 01752-1234

Description

Malecot Nephrostomy Catheter System, Product Number: M001224110, GTIN: 08714729030980 The Malecot and the Stenting Malecot Catheters are designed for percutaneous drainage within the renal collecting system while maintaining external access. The Stenting Malecot Nephrostomy Catheter with its extended tip is also designed for internal drainage from the ureteropelvic junction to the bladder.

Reason

Certain Malecot Nephrostomy Catheters are breaking at the mid-shaft bond during use.

Action

The firm, Boston Scientific, sent an "Urgent Medical Device Product Removal - Immediate Action Required" letter dated 12/12/2017 to its consignees. The letter described the product, problem and actions to be taken. The consignees were instructed to: identify and remove all affected batches from your inventory; immediately discontinue use of and segregate the affected product, and return it to Boston Scientific. Distributors were notified that this removal is to the customer level and asked to notify any customers who received affected product. Consignees were asked to complete and return the Replay Verification Tracking Form (even if you do not have any product to return) via email: [email protected] or Fax to: Field Action Center at 1-866-213-1806. If you have any questions, your local Sales Representative can answer any questions that you may have regarding this removal or contact Vice President , Regulatory Affairs at 508-683-4398 or email: [email protected] or [email protected].

Distribution

Worldwide Distribution: US (nationwide) in the following: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KA, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, including the District of Columbia and Puerto Rico; and countries of: Argentina, Bolivia, Brazil, Canada, Chile, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Mexico, Panama, Peru, Venezuela, Australia, Bangladesh, Botswana, China, Kuwait, Lebanon, Macau, Malaysia, Philippines, Saudi Arabia, Singapore, South Africa, South Korea, Turkey, United Arab Emirates, and Zimbabwe.

Quantity

422