FDA Recall Terminated

Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.

Recall: Z-1098-2013 · Initiated March 6, 2013

Recall

Recall Number
Z-1098-2013
Event Number
64728
Firm
Carefusion Corporation
FEI Number
3003879246
Product Code
FRN
Status
Terminated
Root Cause
Software design
Initiated
March 6, 2013
Posted
April 15, 2013
Terminated
September 2, 2014
Address
3750 Torrey View Ct, San Diego, CA, 92130-2622

Description

Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.

Reason

The recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.

Action

CareFusion sent an Urgent Medical Device Recall Notification letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and action to be taken. Customers are required to confirm receipt of the notification by returning the Recall Response card to Carefusion by postage-paid, self-addressed mail, fax, or email. Carefusion representatives will contact customers by telephone to schedule field remediation. For questions contact CareFusion Support Center at 888-562-6018, Customer Advocacy at 888-812-3266 and/or Technical Support at 888-812-3229.

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Australia and China.

Quantity

9,241 units