Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
Recall
- Recall Number
- Z-1098-2013
- Event Number
- 64728
- Firm
- Carefusion Corporation
- FEI Number
- 3003879246
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- March 6, 2013
- Posted
- April 15, 2013
- Terminated
- September 2, 2014
- Address
- 3750 Torrey View Ct, San Diego, CA, 92130-2622
Description
Alaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.
The recall was initiated because Carefusion has received reports of a communication error on the Alaris PC unit model 8015 with software version 9.12 when the Alaris EtCO2 module or Alaris SpO2 module is attached.
CareFusion sent an Urgent Medical Device Recall Notification letter dated March 6, 2013 to all affected customers. The letter identified the affected product, problem and action to be taken. Customers are required to confirm receipt of the notification by returning the Recall Response card to Carefusion by postage-paid, self-addressed mail, fax, or email. Carefusion representatives will contact customers by telephone to schedule field remediation. For questions contact CareFusion Support Center at 888-562-6018, Customer Advocacy at 888-812-3266 and/or Technical Support at 888-812-3229.
Worldwide Distribution - USA (nationwide) and the countries of Australia and China.
9,241 units