UniCel DxC 600i, Part Numbers: A25638, A25639, Beckman Coulter, Inc.
Recall
- Recall Number
- Z-1098-2008
- Event Number
- 45960
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- April 5, 2007
- Posted
- May 24, 2008
- Terminated
- January 15, 2013
- Address
- 200 S Kraemer Blvd, Brea, CA, 92822-6208
Description
UniCel DxC 600i, Part Numbers: A25638, A25639, Beckman Coulter, Inc.
Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors. -If a stall occurs during a GLUm test near the time of sample inject, results may be affected. -Affected GLUm results may be high or low.
GLUm stir motors greater than one year old were replaced. Replacements began April 5, 2007. Based on the Risk Assessment, BCI decided that Customer Notification is not needed. GLUm stir motors greater than one year old will be replaced on high volume LX/DxC systems by 3rd quarter of 2007. Replacements began April 5, 2007. A new stirrer design is being developed and is expected to be available before 4th quarter of 2007. Implementation Strategy of Stock Control: 1. A Stop Ship was not necessary based on risk assessment. The possibility of stirrer motor failure may increase with time and duty cycles. Systems running a high volume of tests may have a greater likelihood of exhibiting stirrer motor failure with motors greater than one year old. 2. New Customers were placed on a preventative maintenance plan to have their motor replaced at one year cycle. Completed 3. A Technical Update was issued to notify Hotline and Field Service Personnel of the issues.
Worldwide Distribution - USA including Canada.
111 in the US; 19 in Canada