da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,Box)
Recall
- Recall Number
- Z-1096-2026
- Event Number
- 98187
- Firm
- Intuitive Surgical, Inc.
- FEI Number
- 3001675293
- Product Code
- NAY
- Status
- Open, Classified
- Root Cause
- Packaging change control
- Initiated
- December 15, 2025
- Posted
- January 15, 2026
- Address
- 1266 Kifer Rd, Sunnyvale, CA, 94086-5304
Description
da Vinci SP, SP Access Port Kit, Large Incision Reference Numbers:430073 (2.7-4cm,SP1098, Box) & 430075 (2.7-7cm,SP1098,Box)
Access Port System tray may develop cracks potentially resulting in a sterility breach.
On December 15, 2025, Intuitive Surgical Inc. issued a Urgent: Medical Device Recall Notification via mail and email. Intuitive asked consignees to take the following actions: 1. Follow the warning provided in the SP Access Port Kit User Manual Addendum 2. Identify Affected Product: Please inspect the product label to confirm the Part Number and Lot Number. 3. If the product label indicates a Lot Number within the affected range specified, proceed with the following actions 4 and 5 below. 4. Quarantine affected inventory: Immediately segregate and quarantine all affected units to prevent clinical use. 5. Return affected product. 6. Forward this notice to all individuals within your organization who need to be aware. 7. Complete the attached Acknowledgement Form immediately and return it to Intuitive via fax or email as instructed on the form. 8. Please retain a copy of this letter, place a copy with your affected system, and keep the acknowledgement form for your files. 9. Please inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems. 10. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Worldwide - US Nationwide distribution including in the states of Alabama, Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Nevada, New Hampshire, New Jersey, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Poland, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom.
41,526 units