Hitachi Oasis MRI System - C-Spine Coil
Recall
- Recall Number
- Z-1096-2017
- Event Number
- 76056
- Firm
- Hitachi Medical Systems America Inc
- FEI Number
- 1528028
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- January 6, 2017
- Terminated
- November 8, 2018
- Address
- 1959 Summit Commerce Park, Twinsburg, OH, 44087-2371
Description
Hitachi Oasis MRI System - C-Spine Coil
The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.
Hitachi sent an Urgent : Device Correction Notice dated January 3, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm plans to visit each site and replace the cable balun and also confirm receipt of the notification letter. The firm will quarantine returned baluns in their warehouse until they are reworked. For further questions regarding this recall please call (330) 425-1313 ext. 3720.
USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
180 units