FDA Recall Terminated

Hitachi Oasis MRI System - C-Spine Coil

Recall: Z-1096-2017 · Initiated January 6, 2017

Recall

Recall Number
Z-1096-2017
Event Number
76056
Firm
Hitachi Medical Systems America Inc
FEI Number
1528028
Product Code
LNH
Status
Terminated
Root Cause
Component design/selection
Initiated
January 6, 2017
Terminated
November 8, 2018
Address
1959 Summit Commerce Park, Twinsburg, OH, 44087-2371

Description

Hitachi Oasis MRI System - C-Spine Coil

Reason

The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.

Action

Hitachi sent an Urgent : Device Correction Notice dated January 3, 2017, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The firm plans to visit each site and replace the cable balun and also confirm receipt of the notification letter. The firm will quarantine returned baluns in their warehouse until they are reworked. For further questions regarding this recall please call (330) 425-1313 ext. 3720.

Distribution

USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.

Quantity

180 units