FDA Recall Terminated

Invivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (Model IP5) a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professional to remotely monitor the vital signs of patients undergoing MRI procedures. Available monitoring from the IP5 can include ECG, SpO2, non-invasive and invasive blood pressure, CO2, respiration, anesthetic agents, gases, and temperature parameters.

Recall: Z-1096-2014 · Initiated October 3, 2013

Recall

Recall Number
Z-1096-2014
Event Number
67424
Firm
Invivo Corporation
FEI Number
1051786
Product Code
MWI
Status
Terminated
Root Cause
Software design
Initiated
October 3, 2013
Posted
February 26, 2014
Terminated
September 12, 2016
Address
12151 Research Pkwy, Suite 200, Orlando, FL, 32826-3222

Description

Invivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (Model IP5) a display and device controller for Philips MRI patient monitoring systems, is intended for use by trained health care professional to remotely monitor the vital signs of patients undergoing MRI procedures. Available monitoring from the IP5 can include ECG, SpO2, non-invasive and invasive blood pressure, CO2, respiration, anesthetic agents, gases, and temperature parameters.

Reason

Invivo Corporation located at 12151 Research Parkway, Orlando, FL 32826 is recalling their Expression Information Portal (Model IP5). Three different IP5 devices lost audio during use.

Action

Philips Healthcare sent an "URGENT - MEDICAL DEVICE RECALL" letter dated October 3, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to refer to the "Action to be Taken by Customer/User section on the attached pages. A Philips representative will contact the customer regarding their affected devices. All affected devices will have a software update installed in order to correct the problem free of charge. Contact your local Philips representative for further information and support.

Distribution

Worldwide Distribution - USA and the countries of Australia, Austria, Denmark, France, Germany, Netherlands, Poland, Singapore, Sweden, Switzerland, United Kingdom.

Quantity

271