FDA Recall Open, Classified

Plum Duo Infusion System, List Number 40002-04-01

Recall: Z-1094-2026 · Initiated December 15, 2025

Recall

Recall Number
Z-1094-2026
Event Number
98129
Firm
ICU Medical, Inc.
FEI Number
3013319212
Product Code
FRN
Status
Open, Classified
Root Cause
Nonconforming Material/Component
Initiated
December 15, 2025
Posted
January 15, 2026
Address
600 N Field Dr, Lake Forest, IL, 60045-4835

Description

Plum Duo Infusion System, List Number 40002-04-01

Reason

ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.

Action

An URGENT: MEDICAL DEVICE CORRECTION dated 12/15/25 was sent to customers. Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately informed of this notice. If the devices are used at another location, please ensure this communication is delivered to that location. 2. If the primary speaker fails, obtain another pump to continue the infusion. Remove the affected pump from clinical use until the speaker can be replaced. 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. 4. If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. Follow-up Actions by ICU Medical: ICU Medical will address the issue described in this letter through a software update. This update will ensure the backup buzzer will always generate audio for alarms, malfunctions and alerts. ICU Medical will contact you to schedule the software correction for this issue. Device Correction Inquiries https://icumed.custhelp.com/app/market-action For any questions regarding this action

Distribution

Worldwide - US Nationwide distribution.

Quantity

423 units