FDA Recall Terminated

Abbott brand CELL-DYN Emerald, CELL-DYN Emerald CPU Board Hematology System; List Number: 09H39-01; Field Replaceable Unit (FRU) Numbers: 8701686901, 8701686902; Product is distributed by Abbott Diagnostics Division, Santa Clara. CA and manufactured in France for Abbott Diagnostics Division of Abbott Laboratories, Abbott Park, IL Intended use: The CELL-DYN Emerald is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories.

Recall: Z-1094-2011 · Initiated October 15, 2010

Recall

Recall Number
Z-1094-2011
Event Number
57202
Firm
Abbott Laboratories
FEI Number
2919069
Product Code
GKZ
Status
Terminated
Root Cause
Other
Initiated
October 15, 2010
Posted
February 10, 2011
Terminated
June 7, 2011
Address
5440 Patrick Henry Dr, Santa Clara, CA, 95054-1113

Description

Abbott brand CELL-DYN Emerald, CELL-DYN Emerald CPU Board Hematology System; List Number: 09H39-01; Field Replaceable Unit (FRU) Numbers: 8701686901, 8701686902; Product is distributed by Abbott Diagnostics Division, Santa Clara. CA and manufactured in France for Abbott Diagnostics Division of Abbott Laboratories, Abbott Park, IL Intended use: The CELL-DYN Emerald is an automated hematology analyzer designed for in-vitro diagnostic use in clinical laboratories.

Reason

There is the potential for the instrument to not power up due to the flash memory issue. This will result in an error message displayed is "No Memory Available" and has the potential for delay in generating patient results.

Action

The firm, Abbott Laboratories, sent a "Product Correction" letter dated October 15, 2010 to all customers. The letter described the product, problem and action to be taken by the customer. The customers were instructed to install a new printer driver in the software location using the provided instructions and to complete and return the Customer Reply Form via fax 1-800-777-0051 or email to [email protected] even if they no longer have the instrument. If you have any questions regarding this information, U.S. customers should call Customer Support at 1-877-4ABBOTT; customers outside the U.S., please contact your local hematology customer support representative.

Distribution

Worldwide distribution: USA including states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, O, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI; and countries including: Argentina, Australia, Brazil, Canada, Dominican Republic, Germany, Singapore, South Korea, St. Eustalius & St. Martin, Trinidad and Tobago.