FDA Recall Open, Classified

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Recall: Z-1091-2026 · Initiated December 12, 2025

Recall

Recall Number
Z-1091-2026
Event Number
98096
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Software design
Initiated
December 12, 2025
Posted
January 14, 2026

Description

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Reason

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Action

Philips notified consignees on about 12/12/2025 via certified mailing. Consignees were instructed to identify any affected systems on hand, circulate the noification to all users of the system for awareness, notify any recipients if the affected systems have been transferred to another organization, maintain the notification with documentation of the system until Philips completes the correction, and complete and return the response form provided. Philips also provided recommended actions for each issue, if encountered. Philips will address the identified issues by implementing a software update (R2.1.15 or R2.2.15) in all affected systems (FCO72200660 to 2.2.15, FCO72200672 to 2.1.15). Philips expects to have this software update released by Q4 2026 (subject to regulatory clearance). Local Philips representatives will contact each consignee to schedule a visit to install the software updates once available.

Distribution

Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine/United Arab Emirate, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Vietnam, Yemen.

Quantity

10,466 units