FDA Recall Terminated

SPI EASY Abutment. Endosseous dental implant abutment.

Recall: Z-1086-2007 · Initiated April 19, 2007

Recall

Recall Number
Z-1086-2007
Event Number
38267
Firm
Thommen Medical, AG Hauptstrasse, 87 Waldenburg Switzerland
FEI Number
3003836985
Product Code
DZE
Status
Terminated
Root Cause
Other
Initiated
April 19, 2007
Posted
July 24, 2007
Terminated
May 12, 2008

Description

SPI EASY Abutment. Endosseous dental implant abutment.

Reason

The undercut on the SPI EASY abutments which engages the stabilization ring of the implant is missing. Although the abutments can still be placed on the implant by the user the parts do not fulfill the quality expectations of the firm.

Action

Letter 5/14/2007. Customers are requested to check their current inventory for the recall product. If found they are to return the product for replacement. If they have implanted the product they are to check patient x-rays for the defect. If defects are found they are to do a restoration.

Distribution

Nationwide.

Quantity

84