FDA Recall Terminated

Agility. Version 3.0 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.

Recall: Z-1084-2014 · Initiated November 25, 2013

Recall

Recall Number
Z-1084-2014
Event Number
67184
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Terminated
Root Cause
Software design
Initiated
November 25, 2013
Posted
March 6, 2014
Terminated
November 8, 2017
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227

Description

Agility. Version 3.0 of the Integrity interface and control software for the Elekta range of medical digital linear accelerators which is associated with Agility collimators.

Reason

During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

Action

Corrective Action #1 - Release of Important Field Safety Notice 200 01 502 043 - "Recommended quality assurance after a calibration procedure" Released 25 November, 2013. Corrective Action #2 - Permanent Solution - All users of Integrity 3.1 software will be informed or a mandatory upgrade to Integrity 3.2 upon its release which will prevent calibration from being inverted. Target Release for Integrity 3.2 will be available at the end of February 2014.

Distribution

Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AL, AK, AZ, CA, CT, FL, GA, IL, KY, LA, MD, MA, MI, MO, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SD, TN, TX, VT, VA, WA, WV, WI, and WY, and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Czech Republic, Cuba, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan , Panama, Peru, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Ukraine, Venezuela, and Vietnam.

Quantity

300