FDA Recall Terminated

CO2 Incubators (General Purpose Incubator), model numbers 370, 371, 380, 381, 3307, 3308, 3310, 3311, 3540, 3541, 3542, 3543, 3584, 3585, 3598, and 3599. Product Usage: Incubators are designed to maintain an optimal environment for the incubation of tissue and cell samples. These models are designed to maintain temperature, carbon dioxide, and humidity levels.

Recall: Z-1079-2013 · Initiated November 29, 2012

Recall

Recall Number
Z-1079-2013
Event Number
64418
Firm
Thermo Fisher Scientific (Asheville) LLC
FEI Number
1523845
Product Code
JTQ
Status
Terminated
Root Cause
Component design/selection
Initiated
November 29, 2012
Posted
April 10, 2013
Terminated
April 14, 2015
Address
401 Mill Creek Rd, Marietta, OH, 45750-4304

Description

CO2 Incubators (General Purpose Incubator), model numbers 370, 371, 380, 381, 3307, 3308, 3310, 3311, 3540, 3541, 3542, 3543, 3584, 3585, 3598, and 3599. Product Usage: Incubators are designed to maintain an optimal environment for the incubation of tissue and cell samples. These models are designed to maintain temperature, carbon dioxide, and humidity levels.

Reason

Incubators manufactured prior to mid 2009 have coated metal hinges that have reportedly failed in the field.

Action

Thermo Fisher sent an Urgent Field Correction Notice letter dated November 19, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to locate all affected products and determine if it is within the recall. If so, they are to return the Acknowledgement Form in order to get replacement hinges and instructions. Upon receipt of the new hinges they are to replace them according to the instructions.

Distribution

Worldwide Distribution - USA (nationwide) including the states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SN, TN, TX, UT, VA, VT, WA, WI, WV & WY. and the countries: AFGHANISTAN, ARGENTINA, AUSTRALIA, AUSTRIA, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, COSTA RICA, DENMARK, EGYPT, ENGLAND, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, ICELAND, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KOREA, LEBANON, MALAYSIA, MEXICO, NEPAL, NETHERLANDS, NEW ZEALAND, NORWAY, PALESTINE, PEOPLES REPUBLIC OF CHINA, PERU, PHILIPPINES, PORTUGAL, QATAR, REPUBLIC OF PANAMA, ROMANIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED ARAB EMIRATES, VENEZUELA, VIETNAM & WEST INDIES.

Quantity

15,567 incubators