MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Recall: Z-1078-2026 · Initiated December 10, 2025

Recall

Recall Number
Z-1078-2026
Event Number
98116
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
IYE
Status
Open, Classified
Root Cause
Software change control
Initiated
December 10, 2025
Posted
January 13, 2026
Address
400 Perimeter Center Ter NE, Ste 50, Atlanta, GA, 30346-1227

Description

MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Reason

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Action

On 12/10/2025 correction notices were emailed to consignees who were asked to do the following: Radiation therapists must always check the meterset values displayed on the MOSAIQ console with the value displayed on the treatment unit console. If a mismatch occurs, follow Caution 9.6 in the MOSAIQ Instructions for Use. Complete and return the acknowledgement form. Firm is working on the correction and will release a Field Safety Modification with the details as soon as possible. Elekta Regional Contact Details: Region America: [email protected] Tel: +1 855 6935358 https://www.elekta.com Region Asia Pacific: [email protected] Tel: + 852 2891 2208 https://www.elekta.com Japan [email protected] Tel: + 81 3 6722 3800 Fax: +81 3 6436 4231 https://www.elekta.com Region China: [email protected] Tel : + 86 10 8012 5012 https://www.elekta.com Region Europe: [email protected] Tel : + 46 8 587 254 00 https://www.elekta.com Region Turkey, India & Middle East [email protected] Turkey Tel: +90 216 444 6374 India Tel: +1-800-103-7454 Middle East Tel: +00 800 4000 5000 https://www.elekta.com

Distribution

US Nationwide distribution in the state of TX.

Quantity