FDA Recall Terminated

Covidien ChemoPlus Chemo Spill Kit, Nonsterile, SKU DP5108K

Recall: Z-1078-2021 · Initiated January 7, 2021

Recall

Recall Number
Z-1078-2021
Event Number
87145
Firm
Cardinal Health 200, LLC
FEI Number
3001236905
Product Code
LYU
Status
Terminated
Root Cause
Process change control
Initiated
January 7, 2021
Terminated
February 14, 2024
Address
3651 Birchwood Dr, Waukegan, IL, 60085-8337

Description

Covidien ChemoPlus Chemo Spill Kit, Nonsterile, SKU DP5108K

Reason

Gowns included in the kits potentially have rips, holes, tears, incomplete and/or open seams, staining, foreign material or debris on the gowns.

Action

URGENT MEDICAL DEVICE CORRECTION NOTICE notification letters dated 1/7/21 were sent to customers. Actions requested on your part: 1) REVIEW your inventory for affected product. 2) QUARANTINE affected product per Attachment 1 within your inventory 3) AFFIX one WARNING LABEL included with this notification to the front of each ChemoPlus" and ChemoBloc" Chemo Spill and Preparation/Administration kits so that it is clearly visible to clinicians, instructing them to remove the affected gown component prior to use. 4) COMMUNICATE this action with all personnel the need to remove the gowns prior to use. 5) NOTIFY any customers to whom you may have distributed or forwarded affected product about this product action. 6) RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to [email protected], whether or not you have affected product. Please contact the Customer Service group for any questions or to arrange for credit and return of any product: " Hospital800-964-5227 " Federal Government800-444-1166 " Distributor800-635-6021 " All other customers888-444-5440 For questions related to the notification, additional warning labels requests and/or acknowledgement form, please contact the market action team at: [email protected] or call 800-292-9332.

Distribution

Global Distribution. US Nationwide.

Quantity

137,346 kits