FDA Recall Terminated

Sekisui Diagnostics Ammonia L3K Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in plasma

Recall: Z-1078-2017 · Initiated November 29, 2016

Recall

Recall Number
Z-1078-2017
Event Number
75818
Firm
Sekisui Diagnostics P.E.I. Inc.
FEI Number
3002806809
Product Code
JIF
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
November 29, 2016
Posted
January 23, 2017
Terminated
July 6, 2017
Address
70 Watts Ave, Charlottetown Canada

Description

Sekisui Diagnostics Ammonia L3K Assay; Catalog Number: 293-80-91 and 293-10 For the IN VITRO quantitative measurement of ammonia concentration in plasma

Reason

Certain lots are showing an atypical decrease in optical density (OD) over time, which may result in an impact to performance at the high end of the linear range.

Action

Sekisui Diagnostics sent an Important Product Correction Notification letter dated November 29, 2016, to all affected customers. Customers were asked to take the following actions: Customer Instructions: Please stop using and discard any of the Ammonia L3K reagent lots listed above in your inventory; Contact your distributor to discuss replacement product. Distributor Instructions: Please send the "IMPORTANT PRODUCT CORRECTION Customer Notification" provided to all customers that received Sekisui Diagnostics Ammonia LJK from the listed lot numbers; Complete, sign and return the attached CONFIRMATION OF NOTIFICATION form by FAX to the number provided. This will: Indicate receipt of this Field Correction, and confirmation of notification of customers that received impacted customers; Act as confirmation of destruction of affected product in inventory; Allow for issuance of replacement product. Customers with questions were instructed to contact Technical Services at 1-800-565-0265 or email [email protected].

Distribution

Worldwide Distribution - US including DE, KS, FL, TX, CA, MI, TN, MA, ID, NJ, RI, IL and Internationally to Austria, Mexico, Canada, India, South Korea, China, Spain, South Korea, and Sri Lanka

Quantity

774 Liters bulk and 565 kits