DCI Equipment Track Mount Light Assembly, 110v

Recall: Z-1077-2007 · Initiated May 1, 2007

Recall

Recall Number
Z-1077-2007
Event Number
38169
Firm
Dental Equipment LLC
FEI Number
1000523896
Product Code
EAZ
Status
Terminated
Root Cause
Other
Initiated
May 1, 2007
Posted
July 12, 2007
Terminated
July 12, 2007
Address
705 S Springbrook Rd, Bldg B200, Newberg, OR, 97132-7057

Description

DCI Equipment Track Mount Light Assembly, 110v

Reason

Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.

Action

On 5/1/07 the dental office contacted the manufacturer via telephone regarding the incident. On 5/2/07 the firm's service representative visited the dental office and replaced the lights.

Distribution

The lights were distributed to one distributor in NY who in turn distributed the lights to one dental office.

Quantity

4 dual track lights