FDA Recall Terminated

Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.

Recall: Z-1076-2014 · Initiated January 6, 2014

Recall

Recall Number
Z-1076-2014
Event Number
67235
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JIX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 6, 2014
Posted
February 24, 2014
Terminated
May 6, 2015
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL) lots 3FD034 and 3HD012. An in vitro diagnostic multi-analyte calibrator for the calibration of Amylase (AMY), Gamma-Glutamyl Transferase (GGT), Lactate Dehydrogenase (LDH), Lipase (LIPL), and Pseudocholinesterase (PCHE) methods on the Dimension Vista System.

Reason

Complaints were received regarding Quality Control (QC) shifts with Pseudocholinesterase (PCHE) and imprecision during PCHE calibration with the Dimension Vista Enzyme 1 Calibrator (ENZ 1 CAL).

Action

A recall notification letter, dated January 2014, was sent to Users.

Distribution

Nationwide Distribution.

Quantity

3496