FDA Recall Terminated

Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h

Recall: Z-1075-2022 · Initiated July 30, 2019

Recall

Recall Number
Z-1075-2022
Event Number
90104
Firm
Shanghai United Imaging Healthcare Co., Ltd. No. 2258 Chengbei Rd, Jiading Ind. Shanghai China
FEI Number
3011015597
Product Code
KPR
Status
Terminated
Root Cause
Software change control
Initiated
July 30, 2019
Terminated
May 17, 2022

Description

Digital Medical X-ray Imaging System, Model: uDR 596i, uDR 592h

Reason

X-ray imaging system positioning image and protocol label is reversed for Flexion and Extension on C-Spine and L-Spine, this will cause the image to be incorrectly labeled and may cause the patient to have a repeat exposure.

Action

On 07/30/19, Customer Notification Letters started being distributed to customers. Customers were asked to contact the firm if they found an inconsistency between the protocol name and the positioning image. The following phone numbers were provided: (855) 221-1552 (toll free) or (832) 699-2799 (direct). The firm issued Field Correction Instructions to Service Engineers to modify the protocol name of Flexion and Extension on C-Spine & L-Spine with the corrected guidelines. In addition, the firm committed to permanently fix impacted systems with a software update.

Distribution

US Nationwide distribution in the state of TX.

Quantity

4