MininephThl Human lgG4 Kit Product Code: ZK009.LR This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents
Recall
- Recall Number
- Z-1073-2017
- Event Number
- 76041
- Firm
- The Binding Site Group, Ltd.
- FEI Number
- 3002808340
- Product Code
- CFN
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- September 11, 2014
- Posted
- January 23, 2017
- Terminated
- May 3, 2017
- Address
- 8 Calthorpe Road, Birmingham United Kingdom
Description
MininephThl Human lgG4 Kit Product Code: ZK009.LR This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
Binding Site sent an Important Information letter dated September 30, 2014, Customers were instructed to retain the notification for records and additional instructions with the firm stating they were updating the instructions for use. Customers with questions were advised to contact their Binding Site representative or [email protected].
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.