FDA Recall Terminated

VITROS 350 Chemistry System, Catalog Number 6802153, Unique Device Identifier Number 10758750002054; IVD. Product Usage: For in vitro diagnostic use. The VITROS 350 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.

Recall: Z-1072-2016 · Initiated December 21, 2015

Recall

Recall Number
Z-1072-2016
Event Number
72979
Firm
Ortho-Clinical Diagnostics
FEI Number
1000136573
Product Code
JJE
Status
Terminated
Root Cause
Error in labeling
Initiated
December 21, 2015
Posted
March 8, 2016
Terminated
May 30, 2018
Address
100 Indigo Creek Dr, Rochester, NY, 14626-5101

Description

VITROS 350 Chemistry System, Catalog Number 6802153, Unique Device Identifier Number 10758750002054; IVD. Product Usage: For in vitro diagnostic use. The VITROS 350 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.

Reason

Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).

Action

Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION Letter (Ref. CL2015-201, dated December 21, 2015) via FedEx overnight courier and/or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address). The notification letter was sent to all customers who received the affected product. Foreign affiliates were informed by e-mail on 12/20/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Technical Services at 1-800-421-3311.

Distribution

Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela

Quantity

US = 885; Foreign: 3221