FDA Recall Terminated

GE Healthcare - Giraffe and Panda iRes Infant Warmer Product Usage: The infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonates transition to the external environment or to provide a controlled open environment. An optional integrated Sp02 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin and pulse rate.

Recall: Z-1072-2014 · Initiated October 1, 2013

Recall

Recall Number
Z-1072-2014
Event Number
67004
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
BTL
Status
Terminated
Root Cause
Component design/selection
Initiated
October 1, 2013
Posted
February 25, 2014
Terminated
July 22, 2016
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare - Giraffe and Panda iRes Infant Warmer Product Usage: The infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonates transition to the external environment or to provide a controlled open environment. An optional integrated Sp02 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin and pulse rate.

Reason

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

Action

GE Healthcare sent an Urgent Medical Device Correction letter dated November 18, 2013, with a return receipt to all affected customers. The letter identified the products, the problem, and safety instructions to be taken by the customer along with the product correction. It is recommended that all potential users in your facility are made aware of this safety notification and the recommended actions. Safety issues described the description and affected product details, safety instructions, photos showing the back of the resuscitation system where the Air and Oxygen fittings are located and an appendix of the Affected Product Serial Number Definition and Configurations of Fittings. For questions or concerns regarding this notification, please call the phone number listed below: United States: 1-800-345-2700.

Distribution

Worldwide Distribution - USA Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL,GA. HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and the countries of Albania, Austria, Australia, Bahrain, Belarus, Belgium, Canada, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Greece, Guatemala, Germany, Honduras, Hong Kong, Hungry, India, Indonesia, Ireland, Italy, Kazakhstan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Norway, Peru, Poland, Romania, Russia, Saudi Arabia, South Africia, Spain, Sweden, Switzerland Turkey, United Arab, United Kingdom, and Venezuela.

Quantity

12, 373 units total, 8,807 = US, 3,394 = OUS, 172 = unknown