FDA Recall Open, Classified

Allura Xper FD20 Biplane; Model Number: 722013;

Recall: Z-1071-2026 · Initiated December 15, 2025

Recall

Recall Number
Z-1071-2026
Event Number
98109
Firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands
FEI Number
3002807880
Product Code
OWB
Status
Open, Classified
Root Cause
Other
Initiated
December 15, 2025
Posted
January 13, 2026

Description

Allura Xper FD20 Biplane; Model Number: 722013;

Reason

Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

Action

On December 15, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). b. Circulate this letter to all users so that they are aware of the issue. Keep this letter with the documentation of the system until Philips inspects and, where needed, corrects your system. c. Establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. d. If you detect a burning odor, smoke, or fire originating from the R-cabinet in the technical room, immediately switch off the system s main power. Do not operate or use the system further. Follow your institution s emergency procedures. e. In case the affected system has been transferred to another organization, please send a copy of this letter to that organization and inform Philips about this transfer through your local Philips representative. f. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. g. Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, India, Italy, Japan, Kenya, Korea, Republic of, Malaysia, Mexico, Nepal, Netherlands, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Yemen.

Quantity

23 units