VITROS 250 Chemistry Systems, Catalog 8132086, Unique Device Identifier No. 10758750004409, and VITROS 250 Refurbished, Catalog 6801759, Unique Device Identifier No. 10758750001330; IVD. Product Usage: For in vitro diagnostic use. The VITROS 250 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Recall
- Recall Number
- Z-1071-2016
- Event Number
- 72979
- Firm
- Ortho-Clinical Diagnostics
- FEI Number
- 1000136573
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- December 21, 2015
- Posted
- March 8, 2016
- Terminated
- May 30, 2018
- Address
- 100 Indigo Creek Dr, Rochester, NY, 14626-5101
Description
VITROS 250 Chemistry Systems, Catalog 8132086, Unique Device Identifier No. 10758750004409, and VITROS 250 Refurbished, Catalog 6801759, Unique Device Identifier No. 10758750001330; IVD. Product Usage: For in vitro diagnostic use. The VITROS 250 Chemistry System performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Increased U90-382 or 6LU condition codes generated by VITROS 250, 350, 5,1 FS, 4600 and 5600 Chemistry Systems when using Calibrator Kit 9, Lot 954. A trend of complaints regarding customers actions following U90-382 or 6LU condition codes was noted (See RES 72289 - VITROS Calibrator Kit 9, lot 954 recall).
Ortho Clinical Diagnostics sent an URGENT PRODUCT CORRECTION NOTIFICATION Letter (Ref. CL2015-201, dated December 21, 2015) via FedEx overnight courier and/or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address). The notification letter was sent to all customers who received the affected product. Foreign affiliates were informed by e-mail on 12/20/2015 of the issue and instructed to notify their consignees of the issue and required actions. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Technical Services at 1-800-421-3311.
Worldwide Distribution - US Nationwide including Puerto Rico, and the countries of Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela
US: 802; Foreign: 2472