FDA Recall Open, Classified

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Recall: Z-1070-2024 · Initiated January 9, 2024

Recall

Recall Number
Z-1070-2024
Event Number
93806
Firm
Ventec Life Systems, Inc.
FEI Number
3013095415
Product Code
CBK
Status
Open, Classified
Root Cause
Process control
Initiated
January 9, 2024
Posted
February 23, 2024
Address
22002 26th Ave Se, Bothell, WA, 98021-4903

Description

VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used with the VOCSN Unified Respiratory System which provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Reason

The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.

Action

On 12/22/2023, the firm sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION via Direct overnight mail, or FedEx, or email, or hand delivered, to customers informing them that Ventec Life Systems is recalling a specific lot (listed below) of the VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) since the bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Customer are instructed to: 1. Operate the ventilator as defined in the Instructions for Use and promptly respond to all alarms activated by the ventilator. 2. Stop use of the affected Patient Breathing Package and replace it with a non-affected product. 3. Inspect current stock and quarantine/segregate any unused affected products to prevent their use. For questions/assistance - contact Ventec Life Systems directly at 844-MYVOCSN (844-698-6276), Monday through Friday between 8:00 a.m. - 5:00 p.m. PDT.

Distribution

U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan

Quantity

250 breathing circuits