FDA Recall Open, Classified

Sterile Procedural Trays, labeled as the following: a. ORAL SURGERY PACK-LF b. EXTRACTION PACK c. ORAL SURGERY PACK d. DENTAL MINOR PACK e. ORAL

Recall: Z-1070-2023 · Initiated November 17, 2022

Recall

Recall Number
Z-1070-2023
Event Number
91386
Firm
MEDLINE INDUSTRIES, LP Northfield
FEI Number
1417592
Product Code
OFX
Status
Open, Classified
Root Cause
Device Design
Initiated
November 17, 2022
Posted
February 6, 2023
Address
3 Lakes Dr, Northfield, IL, 60093-2753

Description

Sterile Procedural Trays, labeled as the following: a. ORAL SURGERY PACK-LF b. EXTRACTION PACK c. ORAL SURGERY PACK d. DENTAL MINOR PACK e. ORAL

Reason

Sterile blades within sterile kits have the potential for the blade to puncture the outer foil layer of the sterile packaging.

Action

Medline Industries, LP notified consignees via email and first-class mail on 11/15/2022. Consignees were instructed to check stock for affected product and immediately quarantine. For any affected units, customers were instructed to over-label the sterile kits with a warning label indicating "Warning To Avoid Injury, Please Handle Blades with Care." No product is to be returned. Customers were also instructed to complete and return the response form and notify customers if further distributed.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Japan, Mexico, Panama, Turks & Caicos, United Arab Emirates.

Quantity

174 cases (750 units)