IMx Total B-hCG Reagent is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti B-hCG Alkaline Phosphatase Conjungate, one bottle of 4-Methylumbelliferyl Phosphate and one bottle of Specimen Diluent.
Recall
- Recall Number
- Z-1069-2007
- Event Number
- 37920
- Firm
- Abbott Diagnostic International, Ltd. 30 Bo Cruce Davila
- FEI Number
- 2623532
- Product Code
- JHI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 10, 2007
- Posted
- July 12, 2007
- Terminated
- February 14, 2012
- Address
- Barceloneta, PR, 00617-3009
Description
IMx Total B-hCG Reagent is composed of one bottle of Anti B-hCG Coated Microparticles, one bottle of Anti B-hCG Alkaline Phosphatase Conjungate, one bottle of 4-Methylumbelliferyl Phosphate and one bottle of Specimen Diluent.
An increase in background counts elevates the results of low level samples up to approximately 6mIU/mL. This is not enough to shift a negative result to positive on an undiluted sample. If low level samples are diluted then the background effect is magnified potentially yelding false positive results for pregnancy and or Dilution Inaccuracy.
The 4/10/2007 Product Recall Letter was sent to all customers who received the affected lots.
Worldwide, including USA, PR, USVI, CANADA, MEXICO, GUATEMALA, COSTA RICA, COLOMBIA, ECUADOR, PERU, CHILE, URUGUAY, ARGENTINA, BRAZIL, DOMINICAN REPUBLIC, EL SALVADOR, GERMANY, JAPAN, SINGAPORE, HONG KONG, TAIWAN, THAILAND, SOUTH KOREA, AUSTRALIA, NEW ZEALAND, HONDURAS, TRINIDAD & TOBAGO, GRENADA, ST. VINCENT, ST. LUCIA, ANTIGUA, BARBADOS, CURACAO, BRITISH VIRGIN ISLANDS, BAHAMAS, PANAMA, and CAYMAN ISLANDS.
3176 kits