FDA Recall Terminated

Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System

Recall: Z-1068-2014 · Initiated November 14, 2013

Recall

Recall Number
Z-1068-2014
Event Number
66893
Firm
Smith & Nephew Inc.
FEI Number
3006760724
Product Code
OMP
Status
Terminated
Root Cause
Device Design
Initiated
November 14, 2013
Posted
February 21, 2014
Terminated
August 4, 2014
Address
970 Lake Carillon Dr, Ste 110, Saint Petersburg, FL, 33716-1130

Description

Renasys EZ PLUS, a Canister Component of of Renasys EZ Negative Pressure Wound Therapy System

Reason

Modification of the bacterial overflow guard (filter) and related changes to product labeling.

Action

The correction strategy is as follows: " An Urgent Medical Device Correction Notice has been prepared to be distributed to all US consignees via Federal Express delivery for confirmation of the delivery. " A copy of the Urgent Medical Device Correction Notice is to be posted on Smith & Nephew's websites, URL corporate, http://www.smith-nephew.com/news-and-media/, and myrenasys.com, concurrent with the distribution of the Notice to consignees. " US distribution of the original design of Canister Kits codes (66800912, 66800913, 66801066, 66800423 and 66800058) subject to this correction ceased November 1, 2013. " The correction has been implemented on a change being affected basis. The modified design is being distributed to US customers effective November 1, 2013. No product will be removed from the market. " Periodic progress reports to FDA, as required per 21 CFR 806 will be submitted monthly until completion of the correction actions..

Distribution

Worldwide Distribution: USA (nationwide) including states of: AL, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, MT, NC, NH, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, VA, WA, WV, and Puerto Rico; and the countries of: Austria, Australia, Belgium, Switzerland, China, Cyprus, Germany, Denmark, Spain, Finland, France, UK, Greece, Israel, Italy, Malta, Netherlands, Norway, Paraguay, Russia, Saudi Arabia, Sweden, Thailand, Turkey, South Africa, Mexico, and Canada.

Quantity

6,871,375 canisters