FDA Recall
Terminated
TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.
Recall: Z-1067-2013
·
Initiated November 30, 2011
Recall
- Recall Number
- Z-1067-2013
- Event Number
- 64667
- Firm
- Toshiba American Medical Systems Inc
- FEI Number
- 2020563
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- November 30, 2011
- Posted
- April 8, 2013
- Terminated
- June 17, 2013
- Address
- 2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92780-7047
Description
TSX-IOI: AQUILION 32/64; System, X-Ray, Tomography, Computed. Designed to produce cross-sectional images of a human body.
Reason
The firm initiated this recall due to a potential software issue. The DLP (Dose Length Product) value may be displayed incorrectly when the vHP option is used.
Action
Toshiba sent "URGENT: MEDICAL DEVICE CORRECTION" letters to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter contained a return reply form that is to be faxed to TAMS for retention. Contact the firm at (800) 421-1968 for questions relating to this letter.
Distribution
Nationwide Distribution including the states of WA, OH, CA, TX, WV, IL, AZ, CA, WI, ME, WI, NC, FL, MT, and HI.
Quantity
23 units