Fujinon Endoscope Series; Model # EPX-4400 and EPX-4400HD. Digital processing of images from endoscope.
Recall
- Recall Number
- Z-1067-2011
- Event Number
- 57676
- Firm
- Fujinon Inc
- FEI Number
- 2431293
- Status
- Terminated
- Root Cause
- PMA
- Initiated
- January 7, 2011
- Posted
- February 9, 2011
- Terminated
- June 4, 2012
- Address
- 10 Highpoint Dr, Wayne, NJ, 07470
Description
Fujinon Endoscope Series; Model # EPX-4400 and EPX-4400HD. Digital processing of images from endoscope.
The FICE feature installed in video endoscopic processors is being disabled since this feature should have been reviewed under the premarket notification/510k process.
Fujinon sent an Urgent Medical Device Corrective Action Letter, dated 1/7/2011, via overnight mail to all consignees. The letter advises users that the FICE feature should no longer be used and that a Fujinon Field Service Engineer will be coming to physically disable the FICE feature on the processor. Instructions labeled as Attachment B are provided to assist in ceasing use of the FICE function. Attachment A is provided for frequently asked questions. Customers are to complete the Response/Acknowledgement Form and either fax it or email back to the firm. If there are additional questions, customers can call 800-385-4666 or email [email protected].
Nationwide Distribution
783 units