FDA Recall Terminated

Paltop 1.25 Hex Drivers, Long Part Number: 60-70101

Recall: Z-1063-2019 · Initiated February 22, 2019

Recall

Recall Number
Z-1063-2019
Event Number
82374
Firm
Keystone Dental Inc
FEI Number
3005990499
Product Code
DZE
Status
Terminated
Root Cause
Process control
Initiated
February 22, 2019
Terminated
June 19, 2020
Address
154 Middlesex Tpke, Burlington, MA, 01803-4403

Description

Paltop 1.25 Hex Drivers, Long Part Number: 60-70101

Reason

Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature

Action

Keystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete the Reply Communication form.

Distribution

FL, MA, SC, VA

Quantity

3 hex Driver