Paltop 1.25 Hex Drivers, Long Part Number: 60-70101
Recall
- Recall Number
- Z-1063-2019
- Event Number
- 82374
- Firm
- Keystone Dental Inc
- FEI Number
- 3005990499
- Product Code
- DZE
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- February 22, 2019
- Terminated
- June 19, 2020
- Address
- 154 Middlesex Tpke, Burlington, MA, 01803-4403
Description
Paltop 1.25 Hex Drivers, Long Part Number: 60-70101
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
Keystone Dental initiated recall by phone to each one of the impacted customers. The phone call was then followed by a Recall Notification Letter that was mailed via overnight FedEx mail on February 22, 2019. Letters states problem and action to take: Replacement 1.25mm Hex Driver(s) have been provided to you with this letter. The 1.25mm Hex Driver(s) currently in your kit should be discarded. Please refer to the images on page 2 to locate the driver within your kit. This issue is limited to this 1.25mm Hex Driver(s) only and no other product within the kit is affected by this issue. Complete the Reply Communication form.
FL, MA, SC, VA
3 hex Driver