ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
Recall
- Recall Number
- Z-1063-2015
- Event Number
- 70361
- Firm
- Siemens Medical Solutions USA, Inc.
- FEI Number
- 2936884
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Vendor change control
- Initiated
- January 12, 2015
- Posted
- February 7, 2015
- Terminated
- June 10, 2015
- Address
- 685 E Middlefield Rd, Mountain View, CA, 94043-4045
Description
ACUSON S2000 Automated Breast Volume Scanner (ABVS), model number 10434730. The affected sub-components are #10434968 - Boom Welding and 10434969 - Arm Welding. The S2000 ABVS ultrasound System for imaging the breast allows volumetric acquisition of the breast image.
There is a low probability the scanner arm will become completely detached from the scanner column.
Siemens sent a Customer Safety Advisory Notification to all affected sites on January 12, 2015, certified mail, with return receipt. The letter identified the affected product, the issue, and potential risk. Customers were asked to immediately discontinue use of the affected product. A Customer Service Engineer will contact customers to schedule an appointment to make the necessary repairs. Questions should be directed to a local service support person for information regarding timelines and status. Customers are to share the information with all personnel, who need to be aware of the issue.
Worldwide Distribution -- US, including the states of TX and MN, and the territory of P.R.; and, the countries of China, Germany, Russian Fed, Saudi Arabia, Taiwan, France, Thailand, and Italy.
24 units