Symbiq Pump Set, detachable, convertible pin piggyback w/backcheck valve, 3 prepierced reseal ports, Option-Lok, Latex-Free, Non-DEHP, 130 in. (330 cm), 15 drops/mL; a sterile Rx single use device; Hospira Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; product 196750428, list number 19675-28
Recall
- Recall Number
- Z-1063-2010
- Event Number
- 54080
- Firm
- Hospira Inc.
- FEI Number
- 3013319212
- Product Code
- FPA
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 24, 2009
- Posted
- March 9, 2010
- Terminated
- January 29, 2012
- Address
- 375 N Field Dr, Lake Forest, IL, 60045-2513
Description
Symbiq Pump Set, detachable, convertible pin piggyback w/backcheck valve, 3 prepierced reseal ports, Option-Lok, Latex-Free, Non-DEHP, 130 in. (330 cm), 15 drops/mL; a sterile Rx single use device; Hospira Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; product 196750428, list number 19675-28
Company has confirmed reports of backflow of fluid past the backcheck valve on Symbiq administration sets.
Hospira, Inc. issued an "Important Administration Set Information" letter dated November 24, 2009 to their direct accounts, informing them of the affected product. Consignees were instructed to return the defective product to the firm. For further information, contact Hospira Medical Communications at 1-800-615-0187.
Worldwide Distribution -- United States, Australia, Canada, Costa Rica, India, Mexico, Netherlands, Philippines, Singapore and United Arab Emirates.
32,688 sets