Echostar Spica 1.5T MRI system, for magnetic resonance imaging of the body.

Recall: Z-1061-2015 · Initiated January 9, 2015

Recall

Recall Number
Z-1061-2015
Event Number
70238
Firm
Alltech Medical Systems America Inc
FEI Number
3007138967
Product Code
LNH
Status
Terminated
Root Cause
Process control
Initiated
January 9, 2015
Posted
February 18, 2015
Terminated
September 15, 2015
Address
28900 Fountain Pkwy, Solon, OH, 44139-4383

Description

Echostar Spica 1.5T MRI system, for magnetic resonance imaging of the body.

Reason

The MRI system safety feature, which lowers the magnetic field in emergency situations, may not work correctly.

Action

Altech Medical Systems America Inc., notified customers on January 9, 2015, of the problem via telephone. Customers were instructed to immediately cease using the product until the correction can be made. A follow-up letter was sent through the mail, which stated that a service engineer would make a site visit to ramp down the magnet until the issue can be corrected. Customers were asked to complete the attached Acknowledgement and Receipt Form and fax it to (440) 424-2255. For questions contact AMSA at 1-855-200-2672.

Distribution

US Distribution in the state of Maryland.

Quantity

2 units