FDA Recall Terminated

Biomet Offset Tibial Tray Adaptor, 5.0 MM, sterile, Biomet Orthopedics, Inc., Warsaw, IN, REF 141491. The Biomet Offset Tibial Tray Adaptor is a knee prosthesis intended for cemented use in patients with the following indications: painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved, correction of valgus, varus or posttraumatic deformity, correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

Recall: Z-1061-2009 · Initiated February 13, 2009

Recall

Recall Number
Z-1061-2009
Event Number
51136
Firm
Biomet, Inc.
FEI Number
1825034
Product Code
JDW
Status
Terminated
Root Cause
Packaging process control
Initiated
February 13, 2009
Posted
April 7, 2009
Terminated
August 11, 2009
Address
56 E Bell Dr, Warsaw, IN, 46582

Description

Biomet Offset Tibial Tray Adaptor, 5.0 MM, sterile, Biomet Orthopedics, Inc., Warsaw, IN, REF 141491. The Biomet Offset Tibial Tray Adaptor is a knee prosthesis intended for cemented use in patients with the following indications: painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved, correction of valgus, varus or posttraumatic deformity, correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.

Reason

A 2.5mm locking insert was packaged with the 5.0mm Offset Tibial Tray Adaptor instead of the 5.0mm locking insert.

Action

Consignees were notified by "Field Safety Corrective Action" letter dated February 13, 2009 to conduct a recall to the hospital level and to locate and return the recalled product to Biomet Orthopedics. Customers were also instructed to complete the enclosed "Fax Back Response Form" and return it via fax to Biomet Orthopedics at 1-574-372-1683. Direct questions about this recall to 1-800-348-9500 or 1-574-372-3983, Monday through Friday, 8 a.m. to 5 p.m.

Distribution

Worldwide Distribution including Germany, Italy, Netherlands, Spain and the United Kingdom.

Quantity

17