Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.
Recall
- Recall Number
- Z-1060-2019
- Event Number
- 82272
- Firm
- GE Healthcare, LLC
- FEI Number
- 2126677
- Product Code
- IKD
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- February 22, 2019
- Terminated
- August 19, 2020
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.
The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.
Consignees were initially notified via telephone call on about 02/22/2019 and followed up with an "URGENT MEDICAL DEVICE CORECTION" letter. Instructions included to immediately stop use of the affected Integrated ECG cables and quarantine any on hand. GE Healthcare will replace all affected cables at no cost and a GE Healthcare representative will contact each customer to arrange for the collection of the affected cables and provide replacement of new cables. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171.
Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX. Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.
288 devices (23 US, 265 OUS)