FDA Recall Terminated

Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.

Recall: Z-1060-2019 · Initiated February 22, 2019

Recall

Recall Number
Z-1060-2019
Event Number
82272
Firm
GE Healthcare, LLC
FEI Number
2126677
Product Code
IKD
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
February 22, 2019
Terminated
August 19, 2020
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Integrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.

Reason

The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.

Action

Consignees were initially notified via telephone call on about 02/22/2019 and followed up with an "URGENT MEDICAL DEVICE CORECTION" letter. Instructions included to immediately stop use of the affected Integrated ECG cables and quarantine any on hand. GE Healthcare will replace all affected cables at no cost and a GE Healthcare representative will contact each customer to arrange for the collection of the affected cables and provide replacement of new cables. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171.

Distribution

Worldwide Distribution - US Nationwide to AK, KS, MA, MA, ME, NJ, NJ, NM, NY, TX. Distributed worldwide to Chile, Finland, France, Germany, Italy, Korea (Republic Of), Russia, Spain, Sweden.

Quantity

288 devices (23 US, 265 OUS)