FDA Recall Terminated

PROSTIVA RF Model 8929 Hand Piece is the delivery system component of the PROSTIVA RF Therapy System. PROSTIVA RF Therapy Kit 8929, Sterilized using ethylene oxide, Manufacturer Medtronic Inc Minneapolis, MN 55432-5604, Manufactured at Medtronic, Inc Villalba, Puerto Rico. Indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 cm3 and 50 cm3.

Recall: Z-1060-2009 · Initiated June 5, 2008

Recall

Recall Number
Z-1060-2009
Event Number
50856
Firm
Medtronic Neuromodulation
FEI Number
2182207
Product Code
KNS
Status
Terminated
Root Cause
Device Design
Initiated
June 5, 2008
Posted
March 10, 2009
Terminated
May 17, 2012
Address
800 53rd Ave NE PO Box 1250, Minneapolis, MN, 55440-1250

Description

PROSTIVA RF Model 8929 Hand Piece is the delivery system component of the PROSTIVA RF Therapy System. PROSTIVA RF Therapy Kit 8929, Sterilized using ethylene oxide, Manufacturer Medtronic Inc Minneapolis, MN 55432-5604, Manufactured at Medtronic, Inc Villalba, Puerto Rico. Indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 cm3 and 50 cm3.

Reason

Reported incidents of needle retraction failure when using the PROSTIVA RF Model 8929 Hand Piece. The Medtronic Prostiva RF Model 8929 Hand Piece is the delivery system component of the PROSTIVA RF Therapy System.

Action

A Medtronic "Urgent Medical Device Correction" letter addressed to "Healthcare Provider" was sent August 6, 2008. The letter described the issue, product involved and the Prostiva hand piece Disassembly Procedure that should be followed in the event the user is unable to retract the needles during a procedure. The letter also requested the return of the "Medtronic Reply Form/receipt Confirmation" form.

Distribution

Worldwide Distribution -- AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, RI, SC, SD, IN, TX, UT, VA, VT, WA, WI, WY, and WV and countries of Australia, New Zealand, China, Austria, Denmark, Finland, France, Germany, Italy, Kuwait, Lebanon, Norway, Qatar, Russian Federation, Spain, Sweden, Switzerland, and United Kingdom.

Quantity

40,168