FDA Recall Terminated

Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicated for the processing of blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion.

Recall: Z-1059-2019 · Initiated February 11, 2019

Recall

Recall Number
Z-1059-2019
Event Number
82194
Firm
Fenwal Inc
FEI Number
3004548776
Product Code
CAC
Status
Terminated
Root Cause
Software design
Initiated
February 11, 2019
Terminated
June 23, 2020
Address
3 Corporate Dr, Ste 300, Lake Zurich, IL, 60047-8930

Description

Vacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicated for the processing of blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion.

Reason

The optional Vacuum Unit may stop working and display the failure message "Failure vacuum unit" during use.

Action

The firm, Fresenius Kabi, sent the initial distributor a notification on about 02/11/2019 via email. Consignees were notified on about 02/18/2019 via "IMPORTANT - VACUUM UNIT for CATSmart ISSUE" letter dated 2/7/2019. Instructions inducted that the device can still be used and there are not restrictions on the washing process. If an incident is experienced, customers are directed to contact Technical Service at 1-800-937-5060 and to provide the serial number of the device. Fresenius Kabi will be modifying the vacuum pump with a new cable mechanism to provide a secure long-term connection and updated firmware allowing for a consistent signal to the CATSmart device. The modifications will be implemented at customer locations and completed by September 1, 2019. Customers were also instructed to complete and return the Customer Reply form via fax to:1-847-550-0032. If you have any questions, please contact the Fresenius Kabi Technical Service at 800- 937-5060. prompt 2.

Distribution

US Nationwide Distribution to states of: AZ, CA, FL, HI, ID, IL, IN, MD, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OK, OR, PA, TN, TX, VA and PR; and Internationally to Canada.

Quantity

198 devices