LINAC Accelerators: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2, MEVATRON KDS-2, MEVATRON KDS-2, MEVATRON KD-2, MEVATRON K2, MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, MEVATRON M2 6300, MEVATRON MDX-2, MEVATRON MD-2; The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. Product Usage: The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
Recall
- Recall Number
- Z-1059-2015
- Event Number
- 70138
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- December 19, 2014
- Posted
- February 5, 2015
- Terminated
- January 26, 2017
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
LINAC Accelerators: ARTISTE, ONCOR Avant Garde, ONCOR Impression Plus, ONCOR Expression, ONCOR Impression, PRIMUS HI, Mevatron M2/Primus Mid-Energy, MEVATRON PRIMART, MEVATRON KD-2, MEVATRON KDS-2, MEVATRON KDS-2, MEVATRON KD-2, MEVATRON K2, MEVATRON 6323-2, MEVATRON M2 6740, MEVATRON M2 6700, MEVATRON M2 6300, MEVATRON MDX-2, MEVATRON MD-2; The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer. Product Usage: The intended use of the MEVATRON, PRIMUS, ONCOR and ARTISTE MV System linear accelerator system is to deliver x-ray radiation for therapeutic treatment of cancer.
A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
Siemens sent a Customer Advisory Notice Letter TH016/14/S, dated December 19, 2014, to end users that identified the product, problem, and actions to be taken. End users were advised to read the Field Safety Notice and include it with the User's Manual until the hardware update is completed.
Worldwide Distribution - US Nationwide including Canada and Mexico.
310