Randox Blood Gas Control Level 2 -Model BG5002. In-vitro Diagnostic Quality control of Blood Gas analysis This product is intended for in vitro diagnostic use, in the quality control of Blood Gas analysis. Target values and ranges are supplied for the following analytes: Calcium, Chloride, Glucose, Lactate, PCO2, pH, pO2, Potassium, Sodium and Total CO2.
Recall
- Recall Number
- Z-1057-2019
- Event Number
- 82195
- Firm
- Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
- FEI Number
- 1000361607
- Product Code
- JJY
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- February 13, 2019
- Terminated
- April 14, 2020
Description
Randox Blood Gas Control Level 2 -Model BG5002. In-vitro Diagnostic Quality control of Blood Gas analysis This product is intended for in vitro diagnostic use, in the quality control of Blood Gas analysis. Target values and ranges are supplied for the following analytes: Calcium, Chloride, Glucose, Lactate, PCO2, pH, pO2, Potassium, Sodium and Total CO2.
An error in the assignment of the control range pH in Blood Gas Control
The firm, Randox Laboratories Ltd, sent an "Urgent Medical Device Correction " letter dated 2/13/2019 to its customer on 2/13/2019 via email. The letter describes the product, problem and actions to be taken. The customer was instructed todo the following: " Please replace the amended value sheet within the kit of any remaining stock. " Discuss the contents of this notice with your Medical Director. " Randox Laboratories Ltd is not recommending a review of previous results as any issue would be observed at the time of use. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " Complete and return the response form 12187-QA to [email protected] within five working days. If you have any questions or concerns, please contact Randox Technical Services by email at [email protected], or telephone +44 (0)28 9442 2413 or Email: [email protected].
US Distribution to state of: WV
1 kit