FDA Recall Open, Classified

GE Nuclear Medicine 600/800 series systems, specifically: a) BRIVO B615 3/8 inch DHP BJG NR, b) BRIVO 615 3/8 inch, c) BRIVO B615 3/8 inch DHP BJG NR, d) B615 with new rotor, e) BRIVO B615 3/8 inch BJG, f) BRIVO B615 3/8 inch DHP BJG NR, g) B615 with new rotor; Emission Computed Tomography System

Recall: Z-1055-2023 · Initiated December 19, 2022

Recall

Recall Number
Z-1055-2023
Event Number
91456
Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel
FEI Number
3004571294
Product Code
KPS
Status
Open, Classified
Root Cause
Process control
Initiated
December 19, 2022
Posted
February 8, 2023

Description

GE Nuclear Medicine 600/800 series systems, specifically: a) BRIVO B615 3/8 inch DHP BJG NR, b) BRIVO 615 3/8 inch, c) BRIVO B615 3/8 inch DHP BJG NR, d) B615 with new rotor, e) BRIVO B615 3/8 inch BJG, f) BRIVO B615 3/8 inch DHP BJG NR, g) B615 with new rotor; Emission Computed Tomography System

Reason

A mitigation may not be correctly implemented. If this is the case, the detector can fall, leading to life-threatening bodily injury.

Action

GE Healthcare issue an URGENT MEDICAL DEVICE CORRECTION on 12/19/2022 by mail. The notice explained the issue and the risk, and requested the following action by taken: .GE Healthcare asks that you STOP USE of your Nuclear Medicine system, if your system is listed in the Affected Product Details and Appendix of this letter, until an inspection of your system is completed by GE Healthcare and you have been provided with a letter indicating the system can be returned to clinical use. .GE Healthcare will correct all affected products at no cost to you. GE Healthcare will inspect your system to ensure that the appropriate mitigation is correctly implemented to address this potential life-threatening concern. A GE Healthcare representative will contact you to arrange for this correction. .If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, El Salvador, Ethiopia, Finland, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Martinique, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uzbekistan, and Vietnam.

Quantity

7 units