Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL PACK, Medline kit SKU DYNJ54488A
Recall
- Recall Number
- Z-1054-2026
- Event Number
- 98154
- FEI Number
- 1417592
- Product Code
- FTN
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- November 26, 2025
- Posted
- January 12, 2026
- Address
- 3 Lakes Dr, Medline Industries, LP, Northfield, IL, 60093-2753
Description
Medline Kits containing B. Braun IV Administration Sets and Pump Administration Sets, labeled as: DR HOLCOMB FACIAL PACK, Medline kit SKU DYNJ54488A
Medline kits contain B. Braun IV Administration Sets and Pump Administration Sets which were recalled due to customer complaints identifying that the check valve components may become stuck in the open or closed positions.
Medline issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/26/2025 via email and USPS first class mail. The notice explained the issue potential risk to the patient, and referred the consignee to the attached B Braun notice for further instructions. The B Braun notice provide "Interim Measures for Users" to provide for the continued use of the product. For questions, contact the Recall Department at 866-359-1704 or [email protected].
US Nationwide distribution.
16 units