Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969
Recall
- Recall Number
- Z-1053-2008
- Event Number
- 46742
- Firm
- Draeger Medical, Inc.
- FEI Number
- 2517967
- Product Code
- FMZ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- January 31, 2008
- Posted
- March 18, 2008
- Terminated
- May 8, 2008
- Address
- 3135 Quarry Rd, Telford, PA, 18969-1042
Description
Draeger Globe Trotter IC, Catalog Number: MU20509, Draeger Medical, Inc., Telford, PA 18969
Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.
The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or [email protected].
Worldwide Distribution-- USA including states MO, OH, and PR, and countries of Australia, Canada, Columbia, Israel, UAE, France, China, Italy, Denmark, Switzerland, and UK.
4 units