FDA Recall Terminated

AxSYM Troponin-I ADV Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Troponin-I ADV is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of cardiac troponin-I (cTnI) in human serum or plasma on the AxSYM System; list 2J44-22

Recall: Z-1051-2009 · Initiated January 19, 2009

Recall

Recall Number
Z-1051-2009
Event Number
50816
Firm
Abbott Laboratories
FEI Number
1415939
Product Code
MMI
Status
Terminated
Root Cause
Employee error
Initiated
January 19, 2009
Posted
April 15, 2009
Terminated
September 11, 2010
Address
100 Abbott Park Rd, Abbott Park, IL, 60064-3502

Description

AxSYM Troponin-I ADV Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Troponin-I ADV is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of cardiac troponin-I (cTnI) in human serum or plasma on the AxSYM System; list 2J44-22

Reason

Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.

Action

Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.

Distribution

Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.

Quantity

2,314 kits