FDA Recall Terminated

Clinitek 50 Urine Chemistry Analyzer; Product codes 6509C, 6513K, 6514C, 6515C, 6516F and 6518C. [The different product codes are for the various countries where the product is distributed).

Recall: Z-1051-04 · Initiated March 12, 2004

Recall

Recall Number
Z-1051-04
Event Number
29444
Firm
Bayer Healthcare LLC 1884 Miles Ave Buildings 110 (Adlake), 17 & 18 Elkhart IN 46514-2291
FEI Number
1810909
Product Code
KQO
Status
Terminated
Root Cause
Other
Initiated
March 12, 2004
Posted
July 20, 2004
Terminated
September 14, 2004

Description

Clinitek 50 Urine Chemistry Analyzer; Product codes 6509C, 6513K, 6514C, 6515C, 6516F and 6518C. [The different product codes are for the various countries where the product is distributed).

Reason

Some units may only have had 1 read head screw installed instead of the three required by the firm's specifications.

Action

Consignees were notified by recall letter on or about March 12, 2003 requesting customers to return their units for inspection and rework.

Distribution

United States, Australia, Canada, France, Germany, India, Japan, Netherlands, Poland, South Korea, Sweden, Switzerland and United Kingdom.

Quantity

424