NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)
Recall
- Recall Number
- Z-1050-2023
- Event Number
- 91467
- Firm
- NovaSignal Corp.
- FEI Number
- 3013368530
- Product Code
- IYN
- Status
- Open, Classified
- Root Cause
- Nonconforming Material/Component
- Initiated
- December 16, 2022
- Posted
- February 1, 2023
- Address
- 2440 S Sepulveda Blvd, Los Angeles, CA, 90064-1784
Description
NovaGuide Intelligent Ultrasound System, Model: NA-RBTSYS1. Comprised of NovaBot Automated Headset (AKA: NeuralBot, Model: NA-RBT1) and Lucid TCD 2.0 Transcranial Doppler Ultrasound (Model: NA-LUCM1)
Transcranial Doppler/ultrasound system tablet has a faulty hard drive, which will cause the system to freeze and reboot. Initially, the system will reboot to a normal operable state but eventually the system will be rendered inoperable.
On 12/16/22, recall notices were emailed to customers. Customers were told they would be contacted by the recalling firm to facilitate the proper disposition of affected devices. If an unexpected reboot occurs, the recalling firm recommends pausing the use of affected devices until unaffected devices arrive. Complete and return the customer response form. Customers with questions can contact Customer Operations at [email protected]
US Nationwide and Worldwide Distribution: AZ, TX, SC, PA, IL. OUS: CA, SE, NL