LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips
Recall
- Recall Number
- Z-1050-2019
- Event Number
- 82214
- Firm
- LW Scientific, Inc.
- FEI Number
- 3000251573
- Product Code
- JQC
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- February 6, 2019
- Terminated
- May 7, 2021
- Address
- 865 Marathon Pkwy, Lawrenceville, GA, 30046-2857
Description
LW Scientific ZIP IQ Combo Centrifuge, ZIC-CMBO-4RT3 Rev B Product Usage: used as a general purpose centrifuge to spin down blood in hematocrit tubes with additional rotors to spin other fluids in test tubes (15ml and 50ml), microtubes, and PCR strips
The clear, plastic, polycarbonate hematocrit rotor may break apart during use and be ejected through the plastic housing, posing a safety hazard to the technician and potentially damaging the surrounding area.
The recall was initiated by telephone on 02/06/2019. Initial requesting that the consignee immediately cease use of the product. Recall notices were disseminated by telephone and email requesting the return of the product beginning 02/13/2019. The recall is to the user level. The firm has posted a notice on their website stating there is a voluntary recall due to a safety issue and asking users to call if they have a unit purchased after 9/28/18.
Worldwide Distribution - US Nationwide and the countries of Philippines, Honduras, El Salvador, Australia, Guatemala, Canada, Singapore, Australia, Peru, Nicaragua, Trinidad and Tobago, and New Zealand
16 units