AxSYM Myoglobin Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Myoglobin is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of myoglobin in human serum or plasma on the AxSYM System; list 3E43-20
Recall
- Recall Number
- Z-1050-2009
- Event Number
- 50816
- Firm
- Abbott Laboratories
- FEI Number
- 1415939
- Product Code
- DDR
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- January 19, 2009
- Posted
- April 15, 2009
- Terminated
- September 11, 2010
- Address
- 100 Abbott Park Rd, Abbott Park, IL, 60064-3502
Description
AxSYM Myoglobin Reagent Pack, Abbott Laboratories, Abbott Park, IL 60064; AxSYM Myoglobin is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative determination of myoglobin in human serum or plasma on the AxSYM System; list 3E43-20
Defective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
Consignees were sent a Product Correction-Immediate Action Required letter on 1/19/09. The letter informed users of the reported increased frequency of probe crashes when using the affected reagent pack lots and provided instructions on what actions the laboratory can take if they have a defective reagent pack or experience a probe crash with their system. Users need to inspect each reagent pack for broken or improperly sealed caps; manually open each bottle to ensure the flipper bar raises the reagent cap correctly; close each flipper bar prior to loading reagent pack on instrument; and discard any reagent packs that do not open properly. If a probe crash is observed prior to testing samples again, a probe crash recovery must be performed following the procedure in the Operations Manual. A copy of the communication needs to be provided to any laboratory that you may have forwarded affected lots to. Users were requested to return a customer reply letter acknowledging receipt and understanding of the letter. For additional information, contact Abbott Labs at 1-877-4ABBOTT.
Worldwide Distribution -- USA including Puerto Rico and countries of Germany, Mexico, Hong Kong, and Japan.
822 kits