FDA Recall Terminated

Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036, Physio Control, Inc, Redmond, WA 98052

Recall: Z-1050-2008 · Initiated January 8, 2008

Recall

Recall Number
Z-1050-2008
Event Number
46351
Firm
Physio Control, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 8, 2008
Posted
March 29, 2008
Terminated
September 4, 2013
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

Physio-Control LIFEPAK 20 defibrillator/monitor, AC powered external defibrillator with battery backup. Part Numbers: 3202487-073, 3202487-079, 3202488-036, Physio Control, Inc, Redmond, WA 98052

Reason

Inoperable Device - Potential for delay in therapy or prevention of defibrillation therapy due to corrosion of the Printed Circuit Board Assemblies (PCBA).

Action

The firm contacted Customers by phone and by Urgent-Medical Device Recall letters sent via return receipt requested to their consignees on 1/8/08. The letter advised consignees that the Printed Circuit Board Assemblies (PCBA) might not meet quality requirements. This manufacturing deviation could lead to eventual PCBA corrosion that could render the device inoperative and not ready for use in an emergency under battery power. The firm recommended keeping the unit in service until it could be replaced. Representatives will visit each consignee and replace units.

Distribution

Worldwide Distribution --- USA including states of OK, SC, TX, and WA, and country of Germany.

Quantity

103 total devices - 84 domestic, 19 international