FDA Recall Terminated

These models are not available in the US. Guidant CONTAK RENEWAL 4 RF/HE (model H230, H235) Guidant CONTAK RENEWAL 4 RF HE (model H239 ) cardiac resynchronization therapy defibrillator (CRT-D).

Recall: Z-1050-2007 · Initiated March 23, 2007

Recall

Recall Number
Z-1050-2007
Event Number
37769
Firm
Boston Scientific CRM Corp
FEI Number
2124215
Product Code
NIK
Status
Terminated
Root Cause
Other
Initiated
March 23, 2007
Posted
July 11, 2007
Terminated
March 2, 2008
Address
4100 Hamline Ave N, Saint Paul, MN, 55112-5700

Description

These models are not available in the US. Guidant CONTAK RENEWAL 4 RF/HE (model H230, H235) Guidant CONTAK RENEWAL 4 RF HE (model H239 ) cardiac resynchronization therapy defibrillator (CRT-D).

Reason

Boston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.

Action

An Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.

Distribution

Worldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Polynesia, Portugal, Puerto Rico, Qatar, Saint Lucia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan R O C, Thailand, Tunisia, Turkey, U.S. Virgin Islands, United Arab Emirates, United Kingdom, United States, Uruguay, Venezuela, and West Indies.

Quantity

1,217